PSM à 3 filtres absolu HEPA H-14. Dim°int: L.1190 x P.560 x H.650 mm. CytoULTRA 4 Telstar®

Equipé: Fenêtre coulissante pivotante, lampe UV, 1 rob gaz, 1 rob vide, 2 prises élec.
Cet article à été remplacé par une nouvelle référence "1617581"

Référence : 4250058

Unité de vente : Pièce

CytoULTRA® design has been based in the well known and reputated Telstar® BioULTRA Class II cabinet. An additional third HEPA H14 prefiltration stage right below the workbench adapts the CytoULTRA® to the requirements in the manipulation of hazardous substances like the preparation of cytotoxic medicines as prescribed in several european countries recommendations. It is also ideal class II configuration for reinforced biosafety and maintenance cost savings.
CytoULTRA® is certified by TUV according to both European Standard EN 12469-2000 (class II cabinet) and DIN 12980-2005 (specific to cytotoxic cabinets).
• Only the third stage filters are exposed to contaminants and to dust which may require their replacement when clogged. Accessible from the chamber, they are safely removable even without a full cabinet fumigation process.
• V shaped Hepa pre-filters are safe as they fully contain potential contaminants, easily sealed and inserted in trash bag in the chamber, while running in sterile conditions.
• Adjustable feet to adapt floor unevenness.
• The space below the cabinet has been maximized for the operator comfort and allows all working positions.
• A motorized sliding window with a pneumatic gasket (patented) automatically seals the front window and allows to fully isolate the chamber in case of accident or for fumigation purposes.

Full Control Over Expectations :
• Key for turbo exhaust function and menus.
• Four "function" keys allow choosing the options in the different menus.
• Half speed downflow function (interlocked with the light).
• Large graphic LCD screen with comprehensive display of parameters.
• Status LED for: fan, light, UV light, gas valve and socket.
• Two level password protection (operator and service) replaces keyswitch to prevent non authorised use or service.
• Multilingual menu: English, Spanish, German, French.
• Maintenance menu: UV light timer, decontamination process, alarm report, maintenance history, components run time...
• Setup menu: to change or deactivate the password, lock the controls and get the emergency code.
• Dry contact to be used for remote alarm monitoring or connection to BMS.
• Audible and visual alarms with display of alarm conditions :
- High or low laminar downflow.
- Low inflow velocity / exhaust flow rate.
- Front window out of working position.
- UV light on (interlocked with close window position).
- Standby downflow.
- Failure on main control system.
- Fan failure (dual fan safety system).

Options and Accessories :
A complete range of options and accessories allows :
CytoULTRA meeting all needs:
1. Stainless steel IV bar with 6 hooks.
2. Side windows.
3. Plain tray instead of standard divided surface.
4. Canopy connection for a safe connection to a duct.
5. Anti blow-back valve.
6. Additional charcoal exhaust filter to trap bad smell or chemicals.
7. Ultra Logger software enables remote control and PC monitoring.

Technical Data :
• Hepa H14 pre-filter stage with double air filtration.
• The cabinet is only 795 mm depth and 1995 mm height (compatible with all lab door width).
• Slope front improves body position.
• Stainless steel 304L single piece chamber with rounded corners (side windows as an option).
• Independent fans and airspeed control for exhaust flow and downflow (two fans, two sensors).
• Motorized sliding window with inflatable gasket (100% leaktight).
• Visible command/control panel with large LCD.
• Glareless safety glass window without any visual obstacle at the edge.
• Gas spring counter balance the hinged front panel for servicing and window inner side cleaning.
• Close loop active control with thermoanemometric sensor.
• Turbo exhaust function.
• UV timer.
• 4 languages predefined (English, Spanish, French and German).
• Power fails detection and automatic window closing and sealing (battery integrated).
• Tap kits: Gas (includes solenoid valve) and vacuum.

FEATURES :
External dimensions (uncrated) (DxWxH) : 1400 x 795 x 2115 mm
Chamberl dimensions (DxWxH) : 1190 x 560 x 650 mm
External dimensions (crated) : 1480 x 890 x 2290 mm
Weight (crated / uncrated): 340 / 300 kg
Power: 2.1 / 9.1 kW
Electrical supply: 230 V / 50 Hz
Laminar downflow velocity: 0.40 m/s
Downflow: 995 m3/h
Exhaust flow (Ø.200 mm): 540 m3/h
Inflow velocity: 0.60 m/s
Light intensity: > 1000
Noise level: < 58 dBA
Filters HEPA/ULPA H14: > 99.999% (DOP).

Poids 340 Kg
Volume 3,1 m³
Largeur 0,9 m   (90 cm)
Longueur 1,5 m
Hauteur 2,29 m
Les dimensions, le poids et le volume sont bruts (emballage inclus)